ICP-MS
ICP-MS, inductively coupled plasma mass spectrometry, measures many elements at once down to very low levels by turning a sample into ions in a hot argon plasma.
In ICP-MS, a sample is fed into an argon plasma, a glowing gas that typically reaches about 6,000 to 9,000 °C, hot enough to ionise more than 90 per cent of all elements. The ions then pass into a mass spectrometer, which separates them by their mass-to-charge ratio, most commonly with a quadrupole analyser. Liquid samples are sprayed in as a fine aerosol; solids can be sampled with a laser.
The technique measures dozens of elements at once at ultra-trace levels, detecting more than 50 masses in under half a second. Detection limits for many trace metals are in the parts per billion range or lower, and for most elements can reach far lower still. Multi-collector instruments, called MC-ICP-MS, are used to measure isotope ratios precisely for most elements.
With laser ablation, or LA-ICP-MS, a focused, pulsed ultraviolet laser removes material from spots typically 10 to 150 micrometres across, which suits tiny meteorite grains. In a study of cosmic spherules from Widerøefjellet in East Antarctica, LA-ICP-MS with a 50-micrometre spot was used to measure major and trace elements in individual spherules. Analysts watch closely for contamination: in that study, extra lead in the results was put down to contamination during the analysis itself.